High Frequency Electrocautery

Label:Electrocautery

Jan 10, 20244280

High Frequency Electrocautery

Introduction

 

Electrosurgery is used routinely in eye surgery to cut, coagulate, dissect, fulgurate, ablate and shrink tissue. High frequency (100 kilohertz to 5 megahertz), alternating electric current at various voltages (200–10,000 Volts) is passed through tissue to generate heat. An electrosurgical unit (ESU) consists of a generator and a handpiece with one or more electrodes. The device is controlled using a switch on the handpiece or a foot switch.

 

Features

 

Six output mode

 

a) Cut:400WLoad 500Ω

b) Blend 1: 250WLoad 500Ω

c) Blend 2:200W Load 500Ω

d) Spray Coag 120W Load 500Ω

e)Contact Coag∶ 80W Load 500Ω

f) Standard bipolar Coag.70W(Load 100Ω)

 

Power: 110-220V±22V,50Hz/60Hz±1;

Cut: Pure cut, can make easy operation for any organic,the effec is refine cut,no escharosis on the surface.

Blend 1: The effect is cut with coagulation.

Blend 2: The effect is offering a better coagulation during cut.

Spray Coag.: Called spray electrocoagulation,offrs effective non-contact coagulation, Can use this function together with Argon.

Contact Coag.: Called soft electrocoagulation, or called Point coagulation, small ffect area, for a deep coagulation layer;

Standrad Bipolar Coag.: Called rapid Bipolar coagulation, to provide coagulation function rapidly at boot, which can reduce mis-operation rate during operation.